located on the frame—see section A on the illustrations sheet) or separate
exhalation devices to allow continuous flow of air out of the interface. When
the flow generator is turned on and functioning properly, new air from the
flow generator flushes the exhaled air out through the attached exhalation
port. However, when the flow generator is not operating (eg, power failure),
not enough fresh air will be provided through the system, and exhaled air
may be rebreathed. In this situation, remove the Mirage Swift II
immediately.
The exhaust vent flow is lower at lower system pressures (see section
"Technical Specifications" for Pressure-Flow Curve details). If the system
pressure during expiration is too low, there may not be enough exhaust vent
flow to completely flush the air delivery hose of exhaled air before the start
of the next inspiration.
Discontinue using the Mirage Swift II if you frequently experience nasal
irritation, and consult your physician.
The Mirage Swift II must be cleaned and disinfected/sterilized if it is used
between patients. See section "Reprocessing the Mirage Swift II between
Patients" for details.
Do not modify the components of this system. Intentional modification of
any of the components may cause injury or reduce therapy.
If oxygen is used with the flow generator, the oxygen flow must be turned
off when the flow generator is not operating.
Explanation of the Warning: When the flow generator is not in operation,
and the oxygen flow is left on, oxygen delivered into the ventilator tubing
may accumulate within the flow generator enclosure. Oxygen accumulated
in the flow generator enclosure will create a risk of fire. This warning applies
to most types of flow generators.
Please refrain from smoking while oxygen is in use.
Using a mask may cause tooth, gum or jaw soreness or aggravate an
existing dental condition. If symptoms occur, consult your physician or
dentist.
Note: At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen
concentration will vary, depending on the pressure settings, patient
breathing pattern, interface selection and the leak rate.
Parts of the Mirage Swift II