Guidance and Manufacturer's Declaration – Emissions All Equipment and
Systems
This device is intended for use in the electromagnetic environment specified below. The
customer or user of this device should assure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Electrostatic
±6kV contact
Discharge
(ESD)
±8kV air
IEC 61000-4-2
Electrical Fast
±2kV on AC
Transient/burst
Mains
IEC 61000-4-4
±1kV
Differential
Surge
IEC 61000-4-5
±2kV
Common
>95% Dip for
Voltage dips,
0.5 Cycle
short
60% Dip for
interruptions
5 Cycles
and voltage
variations on
30% Dip for
power supply
25 Cycles
input lines
>95% Dip for
IEC 61000-4-11
5 Seconds
Compliance
Electromagnetic Environment -
Level
Guidance
Floors should be wood, concrete or
±6kV contact
ceramic tile. If floors are synthetic,
the relative humidity should be at
±8kV air
least 30%.
Mains power quality should be that
±2kV on AC
of a typical commercial or hospital
Mains
environment.
±1kV
Mains power quality should be that
Differential
of a typical commercial or hospital
±2kV
environment.
Common
>95% Dip for
Mains power quality should be that
0.5 Cycle
of a typical commercial or hospital
60% Dip for
environment. If the user of this
5 Cycles
device requires continued operation
during power mains interruptions, it
30% Dip for
is recommended that the device be
25 Cycles
powered from an uninterruptible
>95% Dip for
power supply or battery.
5 Seconds
EN
SE-DV56
29