SleepCube Enable Menu List

The Enable menu is available in all modes and is a sub-menu that allows you to control
various functions of your SleepCube. It is accessed by pressing the down button while in this
Advanced Menu.
Display Example
Auto-OFF
Enabled
Auto-ON
Enabled
Mask Fit Check
Enabled
Low Backlight
Enabled
Exit Menu
Press 
SE-DV56
16
aDVancED mEnu controlS
Description
This option controls the Auto-OFF feature, which
automatically stops the flow of air after taking off the mask.
The device will require at least 30 seconds to turn OFF
depending on the kind of mask used. When Auto-OFF is
enabled, Auto-ON is also enabled.
This option controls the Auto-ON feature, which
automatically starts the flow of air after breathing once or
twice into the mask. This is always enabled when Auto-OFF is
enabled.
This option controls the mask fit check feature. The air flow
to your mask is constantly measured and if the amount of
airflow exceeds 95 liters/minute for more than 10% of the
time used, a notification will be created. This notification will
be displayed the next time you use the SleepCube.
This is a setting which controls the backlight. Low backlight
Enable will keep the backlight dim during operation. Low
backlight Disabled will turn the backlight OFF during
operation.
Pressing the up key exits the Enable menu.