Caring for Your Test Strips

• Storage condition: 4°C to 40°C (39.2°F to 104°F) for glucose
test strip, below 85% relative humidity. Do NOT freeze.
• Store your test strips in their original vial only. Do not
transfer to another container.
• Store test strip packages in a cool and dry place. Keep away
from direct sunlight and heat.
• After removing a test strip from the vial, immediately close
the vial cap tightly.
• Touch the test strip with clean and dry hands.
• Use each test strip immediately after removing it from the
vial.
• Write the opening date on the strip vial label when you first
open it. Discard remaining test strips after 6 months.
• Do not use test strips beyond the expiry date. This may
cause inaccurate results.
• Do not bend, cut, or alter a test strip in any way.
• Keep the strip vial away from children since the cap and the
test strip may be a choking hazard. If swallowed, promptly
see a doctor for assistance.
For further information, please refer to the test strip package
insert.

Important Control Solution Information

• Use only our control solutions with your device.
• Do not use the control solution beyond the expiry date or
3 months after first opening. Write the opening date on the
control solution vial and discard the remaining solution
after 3 months.
• It is recommended that the control solution test be done
at room temperature 20°C to 25°C (68°F to 77°F). Make
sure your control solution, device, and test strips are at this
specified temperature range before testing.
• Shake the vial before use, discard the first drop of control
solution, and wipe off the dispenser tip to ensure a pure
sample and an accurate result.
• Store the control solution tightly closed at temperatures
between 2°C to 30°C (35.6°F to 86°F). Do NOT freeze.
Reference Values
The device provides you with plasma equivalent blood
glucose results.
Time of day
Fasting and before
meals
2 hours after meals
Source: American Diabetes Association (2012). Clinical Practice
Recommendations. Diabetes Care, 35 (Supplement 1): S1-100.
Please consult your doctor to determine a target range
that works best for you.
EN-18
Normal plasma glucose range
for people without diabetes
(mg/dL)
< 100 mg/dL (5.6 mmol/L)
< 140 mg/dL (7.8 mmol/L)