10. Faults

Problem
No output voltage
Display off
No output voltage
Display on
[OVERLOAD] LED lights
up
Table 1 Faults

11. Repair

Warning:
When carrying out work on the equipment: de-energise the equipment.
-
Repairs must be made by a skilled electrical engineer.
-
When opening the equipment, you may come into contact with parts connected to
-
dangerous voltages.
Simco-Ion recommends that you return the generator if repairs are required.
Request an RMA form by sending an e-mail to service
Pack the bar properly and clearly state the reason for return.
CMMedium_UM_9752099010_NL_D_GB_F_V1_0
Cause
Mains voltage too low or
not present
Fuse at the rear of the unit
is defective
High-voltage generator
defective
A set point has not been
specified through the [ ] / [
] keys or in Remote
Voltage Programming not
via the Sub-D connector)
Remote on/off function has
been switched on, but
signal is not sent to switch
the generator on
Press the [RUN] key to
manually switch the
generator on
High-voltage generator
defective
Sparking or short-circuit in
the equipment
Generator overloaded
Remedy
Restore mains voltage
Replace fuse
Send generator for repair
Specify set point with the
[ ] / [ ] keys or with a signal
0-10 V (when Remote Control
Programming is used) or
disable this option (Sub-D).
Give a signal on the Sub-d
connector to switch the
generator on, or switch off
remote on/off function (remove
solder connection on Sub-d)
Send generator for repair
Check the equipment
Reduce set point
Increase distance between
charging bar/electrode and
material