8. Options

8.1. Summary
Switching on/off (remote on/off) and setting the output voltage are functions that can be
remotely controlled. A 25-pin Sub-D-connector has been fitted at the rear of the unit for this
purpose.
The options can be installed using low voltage / datacable.
Pin Function
1
2
3
4
5
6
7
8
9
10
11
12
13
Sub-D connector layout (rear panel view).
Warning:
When carrying out work on the equipment: de-energise the equipment.
-
-
Work on the equipment must be carried out by a skilled electrical engineer.
Only connect the specified voltages to the control circuits.
-
-
Use only shielded cables and shielded Sub-D connectors for the connections.

8.2. Switching on/off (remote on/off)

8.2.1. Summary
This function switches the output voltage, not the mains voltage. A voltage of 12V (10mA) is
required to switch. This voltage is present on the Sub-D connector, but can also be supplied
externally. Mind the switching speed; the unit needs 60 ms to reach the maximum output
voltage from 0 kV. The time needed to reach 0 kV from the maximum output voltage depends
largely on the connected load.
When the remote on/off is used, a solder connection must be made on the SubD25 connector. If
remote on/off is used, the [RUN] key can be used to activate the output voltage manually for
testing purposes.
CMMedium_UM_9752099010_NL_D_GB_F_V1_0
Remote on/off +12V
n.c.
External set point
n.c.
n.c.
n.c.
+12V internal
n.c.
Open collector overload
n.c.
n.c.
n.c.
n.c.
n.c. not connected
Pin
Function
14
Remote on/off 0V
15
n.c.
16
GND
17
GND
18
Extern set point enable
19
n.c.
20
0V internal
21
n.c.
22
Open emitter overload
23
n.c.
24
GND
25
Remote on/off enable
32