Variations in the operating range
Distortion magnitude
Limit of range
Ambient
0 to 50°C
temperature
Humidity
10 to 90%
no condensation
Power supply
7.5 to 11 V
Frequency
20 to 500 MHz
of field
1 MHz to 1 GHz
1 MHz to 2.5 GHz
Level of
0.1 to 10 V/m
field
0.1 to 100 V/m
0.1 to 200 V/m
Frequency from
Interchangeability
1 MHz to 2.5 GHz
of the sensors
Level from
0.1 to 200 V/m
from 0 to 50°C
from 10 to 90 % HR
power supply
All distortion
from 7,5 V to 9 V
magnitudes
frequency from
1 MHz to 2.5 GHz
Level from
0.1 to 200 V/m
Magnitude distorted
MAX variation
all magnitudes
0.3%/°C
of the reading
± 0.5V/m
per 10°C
all magnitudes
< 0.1 V/m
all magnitudes
0.05 %/V
all magnitudes
± 1 dB
± 1.5 dB
± 2 dB
all magnitudes
± 0.5 V/m
± 1 dB
± 2 dB
all magnitudes
±1 dB
all magnitudes
± 0.5 dB
analog output
± 1 dB of the
from 0 to 20 V/m
display value
Alarm detection
± 0.2 V/m of the
threshold
displayed value
38
Adherence to standards
Class III instrument in accordance with IEC 248.
- Electrostatic discharge (IEC 801-2 and IEC 1000-4-2)
Class of severity :
level 2 (4 kV) no destruction of build components, but change of function that can be reset
by a new instruction.
level 4 (15 kV) non destructive.
- Radiated electrical fields (IEC 801-3 and IEC 1000-4-3)
Class of severity : level 4 (200 V/m)
Power supply
The power supply of the instrument is produced by a 9 V battery type 6 LF 22.
Voltage range ensuring correct operation: 6.5 V to 11 V.
Display of symbol
+
flashing for a battery voltage < 7.5 V (remaining service life
-
approx. 1 hour).
Display of symbol
+
continuously for a battery voltage < 7 V (remaining service life
-
approx. 10 minutes).
Display of the "
" symbol and auto OFF for a battery voltage < 6.5 V (no operation possible,
the battery must be changed).
Average service life: 30 hours in continuous operation without use of the demodulation
function. Each time the instrument is switched ON, the remaining service life (3) as a
percentage of the capacity appears on the display of the C.A 41.
MECHANICAL SPECIFICATIONS
Operating range
Temperature: 0°C to +50°C.
Relative humidity: 10 to 90% RH (no condensation).
Storage range
Temperature: -20°C to +60°C.
Relative humidity: 10 to 95% RH (no condensation).
Adherence to standards (for the measuring instrument)
- Drop resistance: 0.5 m (IEC 68-2-32),
- Shock resistance: 3 shocks of 100g - 11 ms, in the 3 axes (IEC 68-2-27),
- Vibration resistance: 10 cycles from 10 Hz to 55 Hz at 10 g or 0.75 mm in the 3 axes
(IEC 68-2-6),
- Bump resistance: 100 bumps of 10 g in the 3 axes (IEC 68-2-29)
Dimensions and weight
- C.A 41 (without sensor): 212 x 72 x 37 mm - 350 g
- Measurement probe (EF1/EF2): length: 300 mm
diameter: 50 mm
39