Perform self-test

The self-test determines if there are any leaks in the BRAND PLT unit. Two different tests are available.
The Quick Test lasts only about 2 minutes, and should be performed on a daily basis. The Full Test lasts
about 12 minutes. It includes the additional differentiated testing of the internal tanks, and should be
performed on a weekly basis.

Quick test

What to do?
1.
Calling up the
Main Menu
2.
Setting up the
Self Test
3.
Confirming the
Self Test
4.
Selecting the
Quick Test
5.
Confirming
the Quick Test
6.
Seal the system
40
How to do it
The Main Menu is called up
by pressing or turning the
select knob.
Set to "Self Test" by turning
the select knob.
Press the "Execute" button.
The "Self Test" submenu
is called up. The last test
to be performed appears
enframed.
Set to "Quick Test" by turn-
ing the select knob.
Press the "Start/Execute"
button.
The "Start Menu" submenu
is called up.
Press the provided blind
plug firmly into the adapter
(for the multi-channel
adapter, seal all 4 chan-
nels).
Keys to press
Display readout
Pipette Type
Self Test
Switch OFF
Settings...
select
select
Pipette Type
Self Test
Switch OFF
Settings...
select
Quick Test
Full Test
select
Quick Test
Full Test
select
Please insert
blind plug
start
Main Menu
back
back
Main Menu
back
Self Test
back
Self Test
back
Self Test
back