What to do?
7.
Starting the
Full Test
8.
Aerating
the system
9.
Ending the
Self Test
How to do it
Press the 'Start/Execute'
button.
The six components - sen-
sor, pump, valve, filter, mea-
surement chamber, and the
tanks - are tested one after
the next. The progress bar in
the display indicates the test
status. The LED indicator
alternately blinks red and
green during the test.
After completion of the
component tests, the display
shows a prompt to remove
the plug.
Remove the plug and press
the 'Execute' button. The
system is aerated and the
indicator will show the cur-
rent pressure.
After aeration of the sys-
tem, the display changes
automatically and indicates
whether the unit passed the
test. In the case of an error,
the system stops the proce-
dure and indicates an error
code (see page 64), and the
system is aerated.
The test can be repeated by
pressing the 'Start/Execute'
button. The 'Back' button
leads directly to the Main
Menu.
Perform self-test
Keys to press
Display readout
-0.10 hPa 1:sensor
(pressure value is only an
example)
Please remove
blind plug
continue
-20.10 hPa
(pressure value is only an
example)
PASS
restart
green
error: 1XXX
see instruction manual
restart
red
Self Test
back
Self Test
back
Self Test
back
Self Test
back
Self Test
back