Testing Single-Channel Pipettes - BRAND PLT unit Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Example: 100 μl Air-displacement pipette, dynamic test
Before testing set pipette to nominal volume. Mount a new, unused, disposable tip, and insert the pipette
with the tip in the previously selected and attached Adapter 1 (see page 46).
What to do?
1.
Calling up the
Main Menu
2.
Selecting the
pipette type
3.
Confirming the
pipette type
4.
Selecting the
single-channel
pipette
5.
Confirming the
single-channel
pipette
6.
Selecting the
pipette volume

Testing single-channel pipettes

How to do it
The 'Main Menu' is called
up by pressing or turning the
select knob.
Set to 'Pipette Type' by
turning the select knob.
Press the 'Execute' button.
The 'Channels' submenu is
called up.
Set to 'Single-Channel' by
turning the select knob.
Press the 'Execute' button.
The 'Nominal Volume'
submenu is called up.
'100 μl' is selected by turn-
ing the select knob.
Continued on next page
Keys to press
Display readout
Pipette Type
Self Test
Switch OFF
Settings...
select
Pipette Type
Self Test
Switch OFF
Settings...
select
Single Channel
Multichannel
select
Single Channel
Multichannel
select
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
Main Menu
back
Main Menu
back
Channels
back
Channels
back
back
back
47