Relation-
ship of Q
to
L
volume loss
(approx.)
With a leak of a given
magnitude, the volume
loss of the pipette dur-
ing pipetting depends
on various factors,
e.g., the tip opening.
Due to this it can vary
significantly.
(Tolerance for the mea-
sured leak rate: ± 20%)
Pipette tested
Channels
Nominal volume
1-channel
1 μl
1-channel
2 μl
1-channel
2,5 μl
1-channel
5 μl y
1-channel
5 μl
1-channel
10 μl y
1-channel
10 μl c/n
1-channel
20 μl y
1-channel
20 μl c
1-channel
25 μl
1-channel
50 μl
1-channel
100 μl
1-channel
200 μl
1-channel
250 μl
1-channel
300 μl
1-channel
500 μl
1-channel
1000 μl
1-channel
1250 μl
1-channel
1500 μl
1-channel
2000 μl
1-channel
2500 μl
1-channel
5 ml
1-channel
5 ml soft
1-channel
10 ml
1-channel
10 ml soft
4-channel
1 μl
4-channel
2 μl
4-channel
2,5 μl
4-channel
5 μl
4-channel
10 μl
4-channel
20 μl
4-channel
25 μl
4-channel
50 μl
4-channel
100 μl
4-channel
200 μl
4-channel
250 μl
4-channel
300 μl
4-channel
500 μl
4-channel
1000 μl
4-channel
1250 μl
Correlation table
Pipette volume
Leak rate Q
loss (approx.)
L
[hPa*ml/s]
[μl]
4
4
6
12
4
4
3
19
6
16
10
14
13
13
12
10
12
26
36
40
66
52
21
39
20
18
19
29
22
8
10
21
13
13
14
44
47
29
25
25
0.05
0.09
0.13
0.1
0.1
0.1
0.1
0.2
0.2
0.5
0.5
0.8
2
4
4
4
8
16
16
16
40
40
41
60
61
0.1
0.2
0.2
0.2
0.2
0.4
1
1
2
3
8
8
8
16
16
61