Connection J1 Remote I/O Interconnections

Figure 7
All the input/output remote signals to/from the controller must be
connected at P1 mating connector. With the provided J1 mating
connector (shipped with pin 3 shorted with pin 9 and pin 4 shorted
with pin 15) make the connection with AWG 24 (0.25 mm
wire to the pins indicated in the figure to obtain the desired
capability. It is a 15-pins D type connector; the available signals are
detailed in the table, the following paragraphs describe the signal
characteristics and use.
Tab. 3
PIN N.
1
2
3
4
5
6
7
8
9
10
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01

Connection J1 Remote I/O Interconnections

Remote I/ O connector
SIGNAL
START/STOP (+)
START/STOP (-)
INTERLOCK (+)
INTERLOCK (-)
SPEED SETTING (+)
SPEED SETTING (-)
SOFT START (+)
SOFT START (-)
+24Vdc
SET POINT (RELAY)
Technical Information
2
) or smaller
INPUT/OUTPUT
IN
IN
IN
IN
IN
IN
IN
IN
OUT
OUT