Technical Specifications; Environmental Conditions - ResMed Swift LT-F Mode D'emploi

Masquer les pouces Voir aussi pour Swift LT-F:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Using a mask may cause tooth, gum or jaw soreness or aggravate an
existing dental condition. If symptoms occur, consult your physician or
dentist.

Technical specifications

Note: The mask system does not contain latex, PVC or DEHP materials.
Pressure–Flow
Curve
60
50
40
30
20
10
0
4
6
Dead Space
Information
Therapy
Pressure
Resistance
Sound
Environmental
Conditions
Gross
Dimensions
Note: The manufacturer reserves the right to change these specifications
without notice.
Note: The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations, the vent
flow rate may vary.
8
10
12
Mask Pressure (cm H
O)
2
Physical dead space is the empty volume of the mask to the
end of the swivel. Using the large cushions this is 94 mL.
4 to 20 cm H
O.
2
Drop in pressure measured (nominal)
at 50 L/min: 0.6 cm H
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in
accordance with ISO 4871. The A-weighted sound power
level of the mask is 25 dBA, with uncertainty 3 dBA. The
A-weighted sound pressure level of the mask at a distance of
1 m is 17 dBA, with uncertainty 3 dBA.
Operating temperature: +5°C to +40°C
Operating humidity: up to 95% non-condensing
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: up to 95% non-condensing.
380 mm (H) x 57 mm (W) x 62 mm (D). Mask fully assembled
with short tube assembly – no headgear.
14
16
18
O and at 100 L/min: 2.0 cm H
2
8
Pressure
(cm H
O)
2
4
8
12
16
20
20
2
Flow
(L/min)
20
29
37
43
49
O.