Table 3
Guidance and manufacturer's declaration-electromagnetic immunity
CF-VCBC11U is intended for use in the electromagnetic environment specified below. The
customer or user of the CF-VCBC11U should assure that is used in such an environment.
Immunity test
IEC60601 test
level
Conducted RF
3 Vrms
IEC61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
effected by absorption and reflection from structures, objects and people.
*
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
2
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Model CF-VCBC11U is
used exceeds the applicable RF compliance level above, the Model CF-VCBC11U should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the Model CF-VCBC11U.
*
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
3
Compliance
Electromagnetic environment-
level
guidance
3 V
Portable and mobile RF
communications equipment should be
used no closer to any part of the Model
3 V/m
CF-VCBC11U, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters.
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey *
be less than the compliance level in
each frequency range. *
Interference may occur in the vicinity of
equipment marked with the following
symbol:
E
, should
2
3
7