Disinfection
NOTE – The disinfection process can only be completed by the manufacturer or by an appropriately trained individual.
Enclosure, power cord
Collection container, filter, tubing and carrying case Do not clean, replace between patients
PROVIDER'S NOTES
No routine calibration or service is required provided the device is used in accordance with the manufacturer's directions. In case of a change of patient, the device must be
reconditioned to protect the user. Reconditioning must only be carried out by the manufacturer or service provider. Between patients:
1.
Visually inspect unit for any damage, missing parts etc.
2.
Ensure that unit and accessories are clean.
3.
Use an independent vacuum gauge to verify the unit provides the proper vacuum level as stated in Specifications.
4.
Discard and replace collection container, filter, tubing, and carrying case between patients.
5.
Wipe the surface using a clean cloth dampened with disinfectant.
TROUBLESHOOTING
DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a
qualified DeVilbiss Healthcare provider or an authorized service center. Opening or tampering with the unit will void warranty.
NOTE
Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO your DeVilbiss
–
Healthcare PROVIDER WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:
PROBLEM
Unit does not turn on when external power is connected. Green external
power light does not illuminate.
Pump runs, but there is no suction.
Low suction.
Unit does not turn on (no external power is connected). 7314P Series Only 1. Check that battery is fully charged and/or charge battery.
Battery will not charge (external power and charge indicator lights should
be illuminated during charge mode) 7314P Series Only
SPECIFICATIONS/CLASSIFICATIONS
Size (including container)
Weight (including container)
Typical Operating Sound Level
Electrical Requirements
Vacuum Range
Air Flow @ Pump Inlet
Disposable Collection Container Capacity
Reusable Collection Container Capacity
Warranty
Internal Battery (7314P Series Only)
Approvals
Environmental Conditions
Operating Temperature Range
Operating Relative Humidity
Operating Atmospheric Pressure
Storage & Transport Temperature Range
Storage & Transport Relative Humidity
Storage & Transport Atmospheric Pressure
SE-7314-1
Recommended disinfection
interval
Between patients
ACTION
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded.
4. Check for leaks or cracks in collection container assembly.
1. Use vacuum regulator knob to increase suction level.
2. Check system for leaks.
1. Verify that both external power and charge indicator lights illuminate.
2. Check power sources and connections.
3. Ensure wall outlet is live by plugging in a lamp.
8.3 H x 8.0 W x 8.5 D (21.1 cm x 20.3 cm x 21.6 cm) (not including AC to DC adapter)
7314P Series - 6.6 lb. (3 kg) (not including AC to DC adapter)
7314D Series - 4.3 lb. (2.0 kg)
55 dBA
100-240V~, 50/60Hz, 1.2A max
50-550 mm Hg +/- 10%*
27 LPM (free flow) typical (may be less when running from internal battery)* (NOTE– Using Flovac Liners may impair the
device's performance.)
800 ml (cc)
1200 ml (cc)
Two-years limited, excluding internal battery (7314P series only) and collection container
90-day
IEC 60601-1:2005+AMD1:2012; EN 60601-1-2: 2015; ISO 10079-1:2015; IEC 60601-1-2:2014; IEC 60529:1989+AMD1:1999
+AMD2:2013; IEC 60601-1-6:2010; IEC 60601-1-11:2015;IP22
CONFORMS TO AAMI STD. ES 60601-1, AAMI STD. HA 60601-1-11, IEC STD. 60601-1-6, ISO STD. 10079-1
CERTIFIED TO CSA STD. C22.2 NO. 60601-1, NO. 60601-1-11, IEC STD. 60601-1-6, ISO STD. 10079-1
7314 Series complies with IEC 60601-1 3rd Edition
32°F (0°C) - 104°F (40°C)
0-95%
10.2 psi (70 kPa) - 15.4 psi (106 kPa)
-40°F (-40°C) - 158°F (70°C)
0-95%
7.3 psi (50 kPa) - 15.4 psi (106 kPa)
Number of
disinfection cycles
20
N/A
; 12V
; 33 W Max
Compatible disinfection method
1:5 chlorine bleach (5.25%) and water
solution, Mikrobac forte, Terralin Protect
N/A
EN - 9