Electromagnetic Compatibility; Electromagnetic Immunity; Electromagnetic Emissions - Pressalit MSCT 1 Manuel D'utilisation Et D'entretien

Chariot de douche et de change
Masquer les pouces Voir aussi pour MSCT 1:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 71

Electromagnetic compatibility

Electromagnetic immunity

Test standard
EN60601-1-2
and EN17966
(Home healthcare
environment)
EN60601-1-2
(Proximity fields
from RF wireless
communications
equipment)
IEC 61000-4-4
(Electrical fast
transients /
bursts)
Conducted
disturbances
induced by RF
Electrostatic
discharge
* If a statically charged person discharges on or near the shower trolley control
box, a fault may occur in the lifting columns, and a reset must be performed . See
"Resetting" .

Electromagnetic emissions

Test standard
60601-1-2 ch . 7 .1 .7 -
CISPR 14
Compliance
level
=> 10V/m
2,7GHz
=> 10V/m
2,7GHz
Pulse
subjection, 6V
Test level / details
30MHz – 1GHz
Test level /
details
15V/m
Frequency range:
80MHz- 6GHz
1% step size
15V/m
Frequency range:
385 - 5785 MHz
Quick pulse simulation Passed
Pulse
subjection, 10V
Statically charged
person discharge on
unit, 8kV and 15kV
Test results
Passed
Test
results
Passed
Passed
Passed
en
Passed (*)
25

Publicité

Table des Matières
loading

Table des Matières