Tilt function

The shower change trolley can be tilted approximately 3 degrees to allow the
water to drain when using the optional shower mattress .
The tilt function is controlled by the wired remote control unit .
en
It is still possible to raise and lower the unit when it is in tilted position .

Resetting

If the lifting function does not work properly or if the two lifting columns are out
of sync, try resetting the lifting system .
18
Press and hold the Tilt up button to tilt the unit .
The motion stops when the unit is tilted approx . 3
degrees, or when the tilt up button is released .
Press and hold the Tilt down button to level the
unit . The motion stops when the unit is level, or
when the tilt down button is released .
Press and hold the up button and the down button
simultaneously .
The columns will move down after a few seconds
while a beep tone is heard .
Keep holding both buttons until the beep tone
disappears .
The lifting columns are now reset .