Safety installations according to IEC 61010-2-010:
Excess temperature protection
Classification according to DIN 12876-1
Alarm message
Environmental conditions according to EN 61 010, part 1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C (for storage and transportation)
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuation of ±10 % are permissible.
Protection class according to EN 60 529
The unit corresponds to Class
Overvoltage category
Pollution degree
Caution:
The unit is not for use in explosive environment
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).
105 °C - fixed value
class I
optical + audible (permanent)
IP31
I
II
2
Water Baths