9. Cleaning and storing the instrument
• Clean your device and cuff carefully only with a slightly mois-
tened cloth.
• The frequency of cleaning depends on the contamination
level of the device. Clean the blood pressure device and cuff
as soon as any dirt appears on the device.
• Do not use detergents or solvents.
• On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
10. Rectifying faults
Error messages can occur when
• measurement error (
appears in the display),
• you move or talk during the measurement (
display),
• the cuff tube is not properly inserted (
play),
• inflation takes longer than 15 seconds (
play),
• the inflation pressure is higher than 300 mmHg (
in the display),
• an error occurs when storing the measured values (
pears in the display),
• exceeding measurement range (
• the data could not be sent to the PC (
display).
In the about cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do not
move or talk. Put the batteries back in if necessary, or else
replace them.
appears in the
appears in the dis-
appears in the dis-
appears
ap-
appears in the display),
appears in the
11. Specifications
Model no.
Measurement
method
Measurement range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
and transport condi-
tions
Power supply
Battery life
29
BM 75
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Cuff pressure 0 – 300 mmHg,
systolic 30 – 260 mmHg,
diastolic 30 – 260 mmHg,
Pulse 40 –199 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 175 mm x W 117 mm x H 50 mm
Approx. 478 g (without batteries)
24 to 36 cm
+10 °C to + 40 °C, 15 % – 85 % relative
air humidity (non-condensing)
- 10 °C to + 60 °C, 10 % – 90 % relative
air humidity, 700 –1060 hPa ambient
pressure
4 x 1,5 V
AAA batteries
For approx. 120 measurements, de-
pending on the blood pressure level
and/or pump pressure