Anti-Tip Tubes; Seat - Zippie Simba Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51

Anti-tip tubes

Sunrise Medical recommends anti-tip tubes for all chairs. When fitting anti-tip
tubes, use a torque of 16 Nm (Fig. 5.23)
1. Slotting the anti-tip tubes into the clamp:
• Press the rear button on the anti-tip tube on the anti-tip tube adaptor, so
that both release pins are drawn inwards.
• Slot the anti-tip tubes (1) into the anti-tip tube adaptor.
• Turn the anti-tip tubes downwards until the release pin locks into the clamp.
• Fit the second anti-tip tube in the same way.
2. Adjusting the anti-tip tubes:
To achieve the correct ground clearance of approx. 1" to 2" (2.5 cm to 5.0
cm), the anti-tip tubes must be raised or lowered.
• Press the anti-tip tube release button, so that both release pins are drawn
inwards.
• Move the inner tube up or down to slot into the height holes provided.
• Release the button.
• Fit the second anti-tip tube in the same way.
• Both wheels should be at the same height.
DANGER!
If the anti-tip tubes are not fitted, or have been fitted incorrectly, there is a risk
of tipping over and of injury.

Seat

Seat Sling:
Remove the screws on the left side of the sling.
Adjust the VELCRO® material to increase the tension of the seat sling.
Tighten up the screws again.
If the screws are difficult to fit back in place, try and locate the holes with
a sharp object. Ensure also that the plastic base is in the correct position
before the screws are tightened again, (5.24).