Fault inspection during ClO
Fault symptom
Measured value too
small
Measured value too
small
417101816_Anhang09_ClO2.doc
2
Cause / fault
Membrane not fitted properly to shaft Screw membrane cap on tighter
Tip of electrode dirty
ClO
Measuring Module
2
-measurement
Rectification
Clean tip of electrode