NOTE
Through selecting the module selection function one returns to configuration module
selection.
The first page of the ClO
4.2

Measuring parameters

The main menu for adjusting the Multronic is accessed through operating the ↑ key in
the measuring mask.
Main menu
The measurement parameters are selected through placing the cursor onto the selected
item and operating the ENTER key.
Measurement
parameter module
selection
(not available at
Multronic OC!)
The ClO
-measuring module that is to be parametered can now be selected. To do so,
2
move the cursor onto the corresponding selected item and operate the ENTER key.
The following measurement parameters can be adjusted for the ClO
Measurement
parameter (ClO
The measuring range function is used to changes the measuring range.
Value range: 0..2.00 mg/l
The settings for the current signal function are distributed on two lines.
The lower limit of the current signal can be adjusted in the first line.
The minimum output current can be adjusted between 0 mA and 4 mA in the first
position. The second position (activated with →-key from the first position) is used
specify the minimum output current that is allocated to the measured value.
The second line is used to specify the measured value allocation for the maximum
output current (20 mA).
Measuring range
0..2 mg/l
417101816_Anhang09_ClO2.doc
The time periods for the controllers must be within realistic limits.
min. controller on-time < controller cycle time < max. metering time
-module configuration menu is opened through selecting page 1.
2
)
2
Value range
0.00 mg/l to 2.00 mg/l
ClO
Measuring Module
2
- 28 -
-measuring module.
2
Rev.: 05-07/10