It must be ensured that the selected measured value for the maximum output current is
larger than the measured value for the minimum output current. If the current measured
value falls below the measured value specified for the minimum output current, the
output current stays at this minimum. The same applies if the current measured value
rises above the allocation for the maximum output current.
The indication of mutual conductance is for information purposes only. This value cannot
be changed. The mutual conductance results from the calibration of the corresponding
ClO
-measuring cell.
2
The module selection function can be used to return to the measurement parameter
module selection.
4.3
Controller parameters
The main menu for adjusting the Multronic can be accessed through operating the ↑-key
in the measuring mask.
Main menu
The controller parameters function can be selected through placing the cursor onto the
selected item and operating the ENTER key.
Controller parameter
module selection
(not available at
Multronic OC!)
The ClO
-measuring module that is to be parametered can now be selected. To do so,
2
move the cursor onto the corresponding selected item and operate the ENTER key.
The adjustment of the ClO
The display of the first page depends on the controller type setting in the module
configuration menu.
The following settings apply to the signal device:
ClO
-controller
2
parameters page 1
(signal device)
The set point (W) function is used to set the nominal value within the measuring range.
Measuring range
0..2 mg/l
The switch difference (XSD) function is used to specify the switch hysteresis around the
set point. The stated percentage refers to the nominal value.
Value range: 0.0 % to 30.0 %
417101816_Anhang09_ClO2.doc
ClO
-controller parameters is carried out on two pages.
2
Value range
0.00 mg/l to 2.00 mg/l
- 29 -
Measuring Module
2
Rev.: 05-07/10