CareFusion MicroRPM Manuel D'utilisation page 25

Analyseur de pression respiratoire
Masquer les pouces Voir aussi pour MicroRPM:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The MicroRPM is intended for use in the electromagnetic environment
specified below. The customer or the user of the MicroRPM should assure
that it is used in such an environment.
Immunity
IEC
Test
60601
test level
Conducted
3 Vrms
RF
150kHz to
IEC61000-4-
80MHz
6
3 V/m
Radiated RF
80MHz to
IEC 61000-
2.5GHz
4-3
Compliance
Electromagnetic Environment
Level
-Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the micro RPM, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
3 V
d = 1.2 x √P
3 V/m
d = 1.2 x √P
800MHz
d = 2.3 x √P
2.5GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from a fixed RF
transmitters as determined by
an electromagnetic site survey,
a should be less than the
compliance level in each
frequency range. b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
22
80MHz to
800MHz to

Publicité

Table des Matières
loading

Table des Matières