CAUTION: Always place oxygen supply tubing
and power cords in a manner that prevents trip
hazard or possible accidental strangulation.
CAUTION: When using the Oxygen Concentrator
in an automobile, boat, or on other DC sources
with the DC power supply, make sure that
the vehicle is started and running before
connecting the Oxygen Concentrator. Failure
to follow these instructions can result in the
power supply not supplying power to the
Oxygen Concentrator.
CAUTION: When the automobile in which you
are using the Oxygen Concentrator unit is
turned off, disconnect and remove the device
from the automobile. Do not store the Oxygen
Concentrator in a very hot or cold automobile
or in other similar, high- or low-temperature
environments.
Note: Portable and mobile RF communications
equipment can effect medical electrical equipment.
CAUTION: DO NOT leave the Oxygen
Concentrator or the Power Supply
plugged into the vehicle if the
ignition is in the OFF position. Doing
so may drain the vehicle's battery.
CAUTION: Operating or storing the Oxygen
Concentrator outside of its normal operating
temperature range and ambient pressure range
(altitude) can affect performance and decrease
battery run time and/or increase battery charge
time. Refer to the Specifications section in this
manual for storage and operating temperature
limits.
CAUTION: For Oxygen Concentrators equipped
with batteries: Store in a cool and dry location
to help ensure the longevity of your battery.
Storing your Oxygen Concentrator for extended
periods of time at high temperatures or with
a fully charged or completely discharged
battery can degrade its overall battery life.
Do not attempt to open the battery; there are
no serviceable parts inside the battery. Keep
batteries away from children.
CAUTION: ONLY USE the Manufacturer's
provided batteries. For proper battery disposal,
contact your Equipment Provider or your local
government agency for disposal requirements.
FreeStyle Comfort
Note: Ensure the cannula is fully inserted and
secure. This ensures that the Oxygen Concentrator
can properly detect inspiration for oxygen delivery.
During inhalation, you should hear or feel oxygen
flow to the prongs of the nasal cannula.
Note: Lithium batteries may permanently
lose capacity when exposed to extremely hot
temperatures with the batteries fully charged
or completely depleted. For extended storage,
it is recommended that batteries be charged
25 to 50% and remain within a temperature
range of -4°F – 68°F (-20°C – 20°C).
Note: The use of some oxygen administration
accessories not specified for use with this
oxygen concentrator may impair its performance.
Recommended accessories are referenced within
this manual.
Note: Do not operate the Oxygen Concentrator
without the air intake filter in place. If a second
filter is provided, insert the replacement filter
before you clean the dirty filter. Clean the dirty
filter in a warm soap and water solution, then dry
thoroughly prior to use.
Note: The Manufacturer does not recommend the
sterilization of this equipment.
Note: There is never a danger of depleting the
oxygen in a room when you use your oxygen
concentrator unit.
WARNING: THE OXYGEN
CONCENTRATOR SHOULD NOT BE
OPERATED OR PLUGGED IN TO
EXTERNAL (AC/DC) POWER WITHOUT
A BATTERY INSTALLED. DO NOT
ATTEMPT TO POWER ON THE UNIT
OR CONNECT IT TO EXTERNAL
POWER WHEN THE BATTERY PACK IS
REMOVED.
ONLY REMOVE THE BATTERY PACK WHEN THE
UNIT IS POWERED OFF AND UNPLUGGED FROM
THE POWER SUPPLY. DO NOT ATTEMPT TO
REMOVE THE BATTERY PACK WHILE THE UNIT
IS STILL ON OR PLUGGED INTO THE POWER
SUPPLY.
Note: If the oxygen concentrator has been stored
for an extended period of time outside of its
normal environmental operating conditions shown
in the table above, it should be allowed to return
to normal operating temperature before being
powered on.