FreeStyle Comfort
Priority # Alarm
Description
7
Low
Concentration
8
System
Malfunction
Flow Rate
(OCD Valve
Failure,
Obstruction
of Gas
Pathway)
9
Low Battery
Warning
10
Breathing
Rate
Exceeded
11
System
Malfunction
Fan Failure
12
System
Malfunction
Low Pressure
13
System
Malfunction
Low Temp
Info
Start up
20 - ENG
PN MN232-C4 J | User Manual
Display
Audible
More Details
Signal
1 beep
If problem persists, contact provider.
every
30 sec,
for up to
5 min
1 beep
If problem persists, contact provider.
every
30 sec
1 beep,
Charge battery.
no
repeat
1 beep
Breathing Rate exceeds 40 bpm.
every
If dose is being falsely triggered, decrease
60 sec
the sensitivity setting. If problem persists,
contact provider.
none
If problem persists, contact provider.
none
If problem persists, contact provider.
none
Allow unit to warm up to room
temperature. If problem persists, contact
provider.
1 info
Screen displays at power-on and you
beep
should hear one beep to confirm buzzer
is functioning
Start up screen display current firmware
revision in bottom right corner.