Field Calibration Test - Kapro Prolaser Multi Beam Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

This laser level left the factory fully calibrated. Kapro
recommends the user check the accuracy of the laser
periodically, especially if the unit falls or is mishandled.
1. Check the height accuracy of the horizontal beam.
2. Check the leveling accuracy of the horizontal beam.
3. Check the leveling accuracy of the vertical beam.
4. Check the accuracy of the plumb line.
1. Checking the Height Accuracy of the Horizontal
beam. (Up and down deviation)
1. Set up the laser on a tripod or on a flat stable surface
between two walls A and B, approximately 5meters/16
feet apart.
2. Position the laser level approximately 0.5 meter from
wall A.
3. Unlock the pendulum and press the button to project the
horizontal and the vertical cross beams towards wall A.
4. Mark the center of the cross beams on the wall as a1
(see figure # 1).
5. Turn the laser 180° towards wall B and mark the center
of the cross beams as b1 on the wall (see figure 2).
14

FIELD CALIBRATION TEST