Knick 830 R Mode D'emploi page 118

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error
Prio-
message
rity
Er.06
1
Er.10
4
ADJ Menu
Error
Prio-
message
rity
Er.07
3
Er.08
3
Passcode Entry
Error
Prio-
message
rity
Er.09
3
E-118
Error
EEPROM error
Input connection does not cor-
respond to parameter setting.
Example: Input across
terminals 1 and 2 (4 - 20 mA)
and parameter setting for
0 - 20 mA
Error
Adjusting current unstable,
drift ≥ ±2 µA/s
Adjusting current invalid,
deviation at 4 mA > ±0.2 mA or
at 20 mA > ±0.4 mA
Error
Invalid passcode
Remedy
Send in the device
for repair
Change input
connection or
parameter settings.
Remedy
Check loop current
and keep it constant
until adjustment is
terminated.
Check loop current
and remedy error.
Remedy
Re-enter passcode.
If passcode is lost,
send in the device.