Electromagnetic
compatibility
Manufacturer's guide and declaration – electromagnetic immunity
This camera is intended for use in the electromagnetic environment specified below.
The user must ensure that it is used in this environment.
IEC 60601
Immunity test
Level of gravity
Electrostatic
± 6 kV in Contact
discharges
± 8 kV in air
IEC 61000-4-2
Rapid transient
± 2 kV for power
peaks
lines
IEC 61000-4-4
± 1 kV for input/
output lines
Electric shocks
Differential mode
IEC 61000-4-5
± 1 kV
Shared mode
± 2 kV
Power dips,
• < 5 % UT –
power failures,
for 10 ms
short power
• 40 % UT –
interruptions
for 100 ms
and variations in
• 70 % UT –
power voltage
for 500 ms
IEC 61000-4-11
• < 5 % UT –
for 5 s
System
3 A/m
frequency
magnetic field
(50/60 Hz)
IEC 61000-4-8
Compliance
Electromagnetic environment - guide
level
± 6 kV
The floor must be wooden, concrete or
± 8 kV
tiled. If the floor is covered with a synthetic
material, the relative
humidity must be at least 30 %.
± 2 kV
The quality of the main power supply
must be that of a commercial or typical
± 1 kV
hospital environment.
± 1 kV
The quality of the electric power supply
must be that of a commercial or typical
± 2 kV
hospital environment.
< 5 % UT
The quality of the main power supply
10 ms
must be that of a commercial or typical
< 40 % UT
hospital environment.
100 ms
If the user of this camera requires it to
< 70 % UT
continue functioning during interruptions
500 ms
of the main power supply, it is advisable
< 5 % UT
for this camera to be powered by an
5 s
inverter or a battery.
3 A/m
The system frequency magnetic field
must be at a level that is characteristic of
location in a commercial or typical
hospital environment.
31