Installing the apparatus
Turn the activation switch
èThe camera takes about 30 seconds to come on. The Comeg logo is displayed on the screen during
this interval.
èThe flashing LED Standby button, and the appearance of "Standby" on the screen
indicate that the camera is properly activated.
Connect the camera head to the control unit via the dedicated connector
Touch the screen, OR press the standby button to access the main menu.
Select a surgical specialty from Laparoscopy, Hysteroscopy, Cystoscopy, ENT, Arthroscopy and
Flexible Endoscopy, by clicking the associated symbol.
èOnce the selection is made, the camera is in surgery mode.
After connecting the light cable and the endoscope, film a white surface and launch the White
Balance by pressing the dedicated button on the touch screen interface, or the camera head's
pre-programmed button.
èWhen the surgical monitor appears indicating "AWB OK" and the chronometer starts, this signals
that the white balance has successfully been set.
èThe chronometer shows the operating time by default (the starting point being when the white
balance is set). However, this chronometer can be set beforehand if necessary, independently of the
white balance.
The video settings have been defined for each surgical specialty. However, they can be modified,
if necessary, during surgery.
è3 of these (Brightness, Sharpness and Red Gain) can be directly accessed from the main screen.
All the others (Gamma, Blue Gain, Color, Red Phase, Gain, Noise Reduction, and Aperture) can be
accessed by clicking the "More Settings" button. This is a drop-down list. To scroll through the
parameters, touch the left section (the parameter headings) with your finger.
Press "Close settings" to exit and return to the main screen.
You can at any time during surgery to return to the factory settings for the surgical specialty by
clicking the "Specialities presets" button.
Press the "STOP" button at the end of the surgery.
You can at this stage:
Return to the main menu by clicking the "Completed" button;
OR
Create a user profile from the settings used during the operation;
OR
Log in if you wish to add this set of settings to your profile.