table 3
For ME EQUIPMENt and ME SYStEMS that are not LIFE-SUPPOrtING
Guidance and manufacturer's declaration - electromagnetic immunity
the KD-595 is intended for use in the electromagnetic environment specified
below. the customer or the user of the KD-595 should assure that it is used
in such an environment.
IMMUNItY test
radiated rf
IEC 61000-4-3
Note 1: at 80 mHz and 800 mHz, the higher frequency range applies.
Note 2: these guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, am and fm radio broadcast and tV
broadcast cannot be predicted theoretically with accuracy. to assess the electromagnetic
environment due to fixed rf transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the KD-595 is used
exceeds the applicable rf compliance level above, the KD-595 should be observed
to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the KD-595.
b over the frequency range 150 kHz to 80 mHz, field strengths should be less than 3 V/m.
IEC 60601
Compliance
test level
level
3 V/m
3 V/m
80 mHz
to 2.5 gHz
ENGLISH
Electromagnetic environment -
guidance
portable and mobile rf
communications equipment should
be used no closer to any part
of the KD-595, including cables,
than the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
recommended separation
distance:
d = 1.2 √ P
80 mHz to 800 mHz
d = 2.3 √ P
800 mHz to 2,5 gHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according
to the transmitter manufacturer
and d is the recommended
separation distance in meters (m).
field strengths from fixed
rf transmitters, as determined
by an electromagnetic site
survey,
should be less than
a
the compliance level in each
frequency range.
Interference
b
may occur in the vicinity
of equipment marked with
the following symbol: