Accessories Supplied; Cleaning Of Accessories - Gima NEW MAMILAT Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic
circuit. When the filter is wet, it's not possible to use the unit therefore the filter should be changed
immediately.
In case of possible contamination or discolouration, change the filter immediately.
Don't use the suction unit without the protection filter fitted. If the suction unit is used in an emergency
or in a patient where the risk of contamination is not know the filter must be changed after each use.
To clean the instrument it is recommended to use a soft and dry cloth with not abrasive and not
solvent cleaning substances.
To disinfect accessories follow directions:
1.
Grip the biberon with your hand and turn the cap in a counter - clockwise direction remove
the cap from the bottle.
2.
Put the two biberons and the two caps into a suitable disinfectant solution
3.
Disinfect accessories with methylated spirits or hypochlorite-based solutions, easily
findable in chemist
The two aspiration tubes can be sterilized on autoclave using a sterilization cycle.
DO NOT LEAVE INTERNAL PARTS COME IN CONTACT WITH LIQUIDS
DO NOT BOIL OR PUT IN AUTOCLAVE THE ACCESSORIES

ACCESSORIES SUPPLIED

DESCRIPTION
Nipple shield
PC milk biberon 250cc
PC milk biberon 125cc
Nebulizer air Tube with connectors
Milk biberon security plug
Antibacterial Filter

CLEANING OF ACCESSORIES

11

Publicité

Table des Matières
loading

Table des Matières