EN
Troubleshooting
Problem
Possible Cause
I have placed my feet
TENS is only available through
on the device but am
the Body Pads. The foot-pads
unable to see the TENS
provide EMS only.
symbol on the display.
When using the Body
The level of stimulation you are
Pads, the stimulation
using may be too high.
feels uncomfortable.
I am unsure of which
therapy mode I am
using when using the
Body Pads.
I suddenly can't feel
You may have pressed the other
stimulation when using
stimulation mode button by
the Body Pads.
accident.
When using the
Your muscles may not yet be
Arthritis Pads, my
strong enough to carry out a full
thigh muscles are
20 minute session.
aching before the end
of the treatment.
When using the
Therapy has started and has
Arthritis Pads, I can't
reached the rest period.
feel the stimulation in
my thigh.
For more information, including Frequently Asked Questions,
please visit www.revitive.com
38
4397_IFU02_16220241.indd 38-39
Solution
Decrease the stimulation level
using the
(-) intensity control on
the device or remote control.
Look at the display on the device. If
the Body Pads are in EMS mode,
you will see the EMS symbol. If
the Body Pads are in
TENS mode,
you will see the
TENS symbol.
Pressing the opposite stimula-
tion mode button (i.e. pressing
the
TENS button when already in
EMS mode or vice versa), will reset
the timer back to 30 minutes. The
intensity will also be reset to zero.
Turn the device off and end the ses-
sion before the end of the 20 min-
utes. On your next session start on
a lower setting (where you can feel
the mild electrical stimulation and
it is comfortable) and reduce the
duration until your muscles have
acclimatised to the stimulation.
You will need to wait for up to 20
seconds before the stimulation will
start again.
Technical Specifications
Name of product
Model
Frequency
Output current
Weight
Dimensions (with IsoRocker® enabled and disabled)
Power consumption
AC adaptor
Battery and battery life
Power source
Input (adaptor used)
Output
The Remote Control replicates the controls found on the device
Operating Frequencies
Operating Range Distance
EN
REVITIVE Arthritis-Knee
2836AB
EMS: 1Hz – 50Hz
TENS: 80Hz - 130Hz
EMS: Max 15mA
TENS: Max 19mA
1.5kg
enabled: 355(W)x355(H)x75(D) mm
disabled: 355(W)x355(H)x90(D) mm
5W
CE Approved
Li-Ion / 500 recharge cycles
100-240V
AC ( ), 50/60Hz, 0.18A
5V (
) DC
,1.0A
38 KH
z
0 to 25m at horizontal
0 to 18m at +/- 30 degree angle
from the horizontal.
39
25/01/2019 14:51