Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Gima ECG600G Manuel De L'utilisateur

Masquer les pouces Voir aussi pour ECG600G:
Table des Matières

Publicité

Guidance and manufacturer's declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The model ECG600G Electrocardiograph(ECG) is intended for use in the electromagnetic
environment
specified
Electrocardiograph(ECG) should assure that it is used in such an environment.
Immunity test
IEC 60601 test
Conducted
3 Vrms
RF
150 kHz to 80
IEC
MHz
61000-4-6
3 V/m
Radiated RF
80 MHz to 2,5
IEC
GHz
61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption
and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
a
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the model ECG600G Electrocardiograph(ECG) is used exceeds the
applicable RF compliance level above, the model ECG600G Electrocardiograph(ECG) should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the model ECG600G Electrocardiograph(ECG).
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
b
below.
The
customer
Compliance level
level
3 V
3 V/m
or
the
user
of
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part
of
the
Electrocardiograph(ECG) including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
3
5 ,
d
[
]
P
=
V
1
3
5 ,
d
[
]
P
=
E
80 MHz to 800 MHz
1
7
d
[
]
P
=
E
800 MHz to 2,5 GHz
1
where p is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
(m).b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,a should be less than the compliance
level in each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following symbol:
41
the
model
ECG600G
model
ECG600G

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

33222

Table des Matières