4. Safety functions
4.1. Emergency stop button
At the controlbox, there is an emergency stop button. After
the emergency stop has been pressed, all electrical operation
will cease to function. Turn the emergency stop a quarter turn
to unlock it again (See figure 7).
The orange LED light on the hand control will flash when
the Lock button has been activated.
4.2. Sleep setting
After around an hour, the stretcher will fall into a 'sleep'
setting and the battery indication LEDs on the manual control will go out. In order to use
the stretcher again, it must be activated (See paragraph 2.2. Activating).
4.2. Emergency lowering button
If the controls stop working and the shower stretcher /
dressing table is in a raised position, it can be lowered using
the emergency lowering button. This button is located to the
left of the emergency stop button (See figure 8).
First check to ensure that the emergency button has not been
pressed!
• Remove the plug from the manual control.
• Press the emergency lower button (to the left of the
emergency stop).
The shower stretcher / dressing table will now return to the lowest position (make sure
no persons or objects are under the stretcher).
Take the less able person from the shower stretcher / dressing table and make sure the
device is not used until it has been repaired. Place the shower stretcher / dressing table
in a safe place and indicate that it is out of order.
Contact the technical service or Lopital's service department.
E-mail: service@lopital.nl
Shower stretcher / Dressing table | Instructions for use | 35
Figure 7 Emergency stop button.
Figure 8 Emergency lowering
button.