• When storing the device, make sure that no heavy objects are
placed on top of it. Remove the batteries. The cuff tube should
not have any sharp kinks.

8. Error messages / troubleshooting

In case of faults, the
Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement (in addition to
the cardiac arrhythmia icon
4. the cuff tube is not inserted properly,
5. the cuff is inflated for longer than 16 seconds.
In the above cases, you must repeat the measurement. Make sure
that the cuff tube is properly inserted and that you do not move or
talk. Re-insert the batteries if necessary, or else replace them.
9. Specifications
Model No.
SBM 15
Method of Measure-
Oscillometric, non-invasive blood pres-
ment
sure measurement on the upper arm
Range of Measure-
Pressure 0 – 300 mmHg,
ment
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
pulse 40 –199 beats/minute
Accuracy of the
systolic ± 3 mmHg /
pressure reading
diastolic ± 3 mmHg /
pulse ± 5 % of the displayed value
Measuring uncertainty
max. permissible standard deviation
according to clinical testing:
systolic: 8 mmHg /diastolic: 8 mmHg
message appears in the display.
appears in the display),
Sets of Memory
Unit Dimension
Unit Weight
Cuff Size
Operating Temperature + 10 °C to + 40 °C, ≤ 85 % RH
Storage- and
transport temperature
Power Supply
,
Battery Life
Accessories
Classification
Protection class
Key to symbols
These specification are subject to change without notice for pur-
pose of improvement.
Adapter
Model No.
FW7333SM/06
Input
100 – 240 V, 50 – 60 Hz
Output
6 V DC, 1150 mA, only in connection with
Sanitas blood pressure monitor
Supplier
Friwo Gerätebau GmbH
17
2 x 60 memory places
L147 x W104 x H66 mm
Approx. 310 g
cuff for upper arm circumference of
23 – 35 cm
- 20 °C to + 70 °C, ≤ 85 % RH
4 x 1,5 V AA batteries (alkaline type LR6)
For approx. 250 measurements,
depending on blood pressure level or
inflation pressure
With storage pouch, instruction manu -
al, 4 pcs "AA" battery
Application part Type BF
internal power supply, IPXO, no AP or
APG, continuous operation
Important! Read the operating
instruction!
Application part Type BF