Quick start guide CellaTemp PA
panel for 2 second. Observe the laser safety instructions. For fo-
cal adjustment, loosen the screw of the measuring head and
move the inner tube to obtain a sharp spot in the measuring area
7. The two-color pyrometer detects the infrared radiation of the tar-
get object on two wavelengths. The temperature is then defined
by the ratio of the two signals. The emissivity, i.e. the radiation
characteristics of the target may change due the nature of the sur-
face or in relation to the temperature, but with simultaneous
changes over both wavelengths there is no influence of the meas-
urement.
Changing the emissivity ratio can compensate for the dif-
ference between measured temperature and true temper-
ature when selective interfering factors or material-related
different emissivities are affecting Lambda 1 and Lambda
2.
8. The emissivity ratio is to be set as following:
► Press[
► Release [
> The current temperature value is displayed and the new emis-
sivity ratio coefficient is stored.
Make sure that the function of the pyrometer is correct. All param-
eters can be set directly at the pyrometer. Record the adjusted pa-
rameters.
9. It is advisable to integrate this pyrometer check into the mainte-
nance plan.
Shipping, packing and disposal
Inspection after shipping
Unpack and inspect the entire shipment immediately upon receipt to
make sure it is complete and undamaged. If the container/package
shows visible signs of damage, please refuse the shipment. If this is not
possible, accept the shipment on the condition that the freight carriers
delivery record is noted with the extent of the damage in order to file a
claim. Should you discover a concealed loss or damage, report it to the
shipper or freight carrier immediately. If the period for filing claims has
expired, you will no longer be able to make any claims for compensation
of damage or loss.
or
until the desired emissivity ratio is shown.
▼]