Unit hours
14. Mask off with/without BEEP setting mode
Press "Mode" at mask off with/without BEEP setting mode, then the de-
vice enters the BEEP setting mode when mask off. There are two choices:
mask off with BEEP or mask off without BEEP. Select proper one as one
like before operation.
Mask off without BEEP
Mask off with BEEP
15. Return to standby mode
Press "Start/Off" button to leave device setting mode and return to
standby mode.
16.Operation with PC
The device setting mode can also be done with PC. Connect the device
and your PC with a USB cable, then start running "CPAP" software on
your PC. This software allows you to operate most of device settings on
PC. When the device is connected to PC, it switches USB mode. The air
output of device will be stopped and LCD of the device will display "USB"
on the screen. Disconnect the USB cable, then the device return to standby
mode again.
USB cable attached
Cleaning and Sterilization
It is recommended that the air tube and mask to be thoroughly cleaned
with hot water after use each time and be cleaned with a mild detergent
after the last treatment of the day. When your physician or respiratory
therapist specifies another cleaning procedure, follow their instructions.
8
Rinsing (After each treatment)
1. Disconnect the air tube and mask.
2. Rinse the mouthpiece and mask with water.
3. Dry them with clean soft towel or let it air dry.
4. Reassemble the product when completely dry and put these parts in a
dry, sealed container.
Disinfections
Please follow the following steps to disinfect your CPAP unless otherwise
specified by your physician. It is recommended that the unit is disinfected
after the last treatment of the day.
1. Use one unit white vinegar to be diluted with three units distilled water.
Make sure mixed solution is enough to submerge the air tube, mouthpiece
and mask.
2. Complete Rinsing Steps 1-3.
3. Wash air tube, mouthpiece and mask in warm water and a mild
detergent. Then wash them in hot tap water.
4. Submerge these parts in the vinegar and water solution for thirty
minutes.
5. Complete Rinsing Steps 3~4.
Filter replacement
1. Do not wash or clean the filter and do not operate without a filter.
2. Replace filter every 30 days or when the filter turns gray.
Replaceable parts and accessories
In order to provide proper treatment and long last device, please use repla-
ceable parts and accessories recommended below when necessary.
Device Specifications
Description
Specification
Power Supply
12V DC
Current Consuming
Max 1.5A
Air Pressure Range
3 cmH
Ramp Time
0 – 60 dk
Air Pressure Calibrate ± 3 cmH
Working Flow
3 ± 9 cmH
Noise
< 9 cmH
Working Humidity
%10 RH - %90 RH
Working Temperature 5 ℃ - 40 ℃ (41 ℉ - 104 ℉ )
Storage Humidity
< %95 RH
Storage Temperature
-20 ℃ - 60 ℃ (-4 ℉ - 140 ℉ )
Dimension (L×W×H)
143 x 121 x 88mm
Weight
400g
Life of Use
2 years
Safety Criterion
CE, FDA
O – 20 cmH
O
2
2
O
2
O'da 0.3m3/h
2
O'da 33 dB (1m away)
2