Installation,commissioning
Ensuring the minimum insertion depth of the measuring
electrode when installing the S022 adapter on a T-fitting, a
tank or directly onto the pipe (Fig. 8 to Fig. 10).
S022
adapter
Fitting
Pipe
Fig. 8 :
Insertion of 22 mm minimum of the electrode on a pH-/
Redox-meter 8202
Type S022
22 mm min.
S022
adapter
Fitting
Pipe
Fig. 9 :
Insertion of 22 mm minimum of electrode C = 0.1 or 0.01
on an 8222 conductivity meter
S022
adapter
Fitting
Pipe
Fig. 10 : Insertion of 19 mm minimum of the electrode on an 8222
conductivity meter
English
22 mm min.
19 mm min.
21