Technical Specifications - ResMed Mirage Quattro Mode D'emploi

Masquer les pouces Voir aussi pour Mirage Quattro:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27

Technical Specifications

Note: The mask system does not contain latex, PVC or DEHP materials.
Pressure–Flow Curve
70.0
60.0
50.0
40.0
30.0
20.0
10.0
0.0
2
4
6
8
10
Mask Pressure (cm H
Note: As a result of manufacturing variations, the vent flow rate may vary.
Dead Space Information
Dead space: the empty volume of the mask up to the swivel.
The dead space of the mask varies according to cushion size. It is 242 mL for the large
size.
Standards
CE designation in accordance with EC directive 93/42/EEC, class IIa.
ISO 14971, ISO 13485, ISO 10993-1 and ISO 17510-2.
Therapy Pressure
4 to 20 cm H
O
2
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.5 cm H
at 100 L/min: 1.6 cm H
Sound
The A-weighted sound power of the mask is 28 dBA.
Environmental Conditions
Operating temperature: +5°C to +40°C
Operating humidity: 10%–95% relative humidity non-condensing
Storage and transport: -20°C to +60°C
Storage and transport humidity: 10%–95% relative humidity non-condensing
12
14
16
18
20
22
24
26
O)
2
O
2
O
2
Pressure
(cm H
O)
2
4
8
12
16
20
28
30
7
Flow
(± 6.0 L/min)
22
32
41
48
54

Publicité

Table des Matières
loading

Table des Matières