2 Unpacking and Installation
2.1 Unpacking and Inspection
Check box and contents for damage
When the instrument is received please check the shipping box for rough handling and damage. If the box is
damaged it should be kept until the contents of the shipment have been checked mechanically and
electrically. If the instrument is faulty please contact your local distributor. Keep the shipping material for the
carrier's inspection and insurance claim.
Keep carton for future shipment
The Air Fx Air Irrigator comes in its own shipping carton, which is specially designed for the Air Fx Air
Irrigator. Please keep this carton. It will be needed if the instrument has to be returned for service.
If service is required please contact your local distributor.
Reporting Imperfections
Inspect before connection
Prior to connecting the product it should once more be inspected for damage. All of the cabinet and the
accessories should be checked visually for scratches and missing parts.
Report immediately any faults
Any missing part or malfunction should be reported immediately to the supplier of the instrument together
with the invoice, serial number, and a detailed report of the problem. In the back of this manual you will find a
"Return Report" where you can describe the problem.
Please use "Return Report"
Please realise that if the service engineer does not know what problem to look for he may not find it, so using
the Return Report will be of great help to us and is your best guarantee that the correction of the problem will
be to your satisfaction.
2.2 Storage
Environmental Conditions
IEC 60601-1 Standards Compliance
Class I device for protection against electric shock
Type B Applied Part for degree of protection against electric shock
IPX0 rating for degree of protection against the ingress of water
(i.e. the system can be damaged if any water is spilled on the electronic equipment)
Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and
needs to be installed and put into service according to the EMC information provided.
Portable and mobile radio frequency (RF) communications equipment (e.g. cell phones, personal
data assistants, etc.) can affect medical electrical equipment. This equipment should not be used at
close distances to the equipment.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment.
The Air Fx may not be used in rooms where there are explosion risks. The
equipment is not suitable for use in the presence of flammable anesthetic mixtures
with air or oxygen or nitrous oxide.
To avoid the risk of electrical shock, the Air Fx must only be connected to a supply
mains with protective earth.