7 Repair
Regularly check the ilter system for leaks. Regularly check the hoses for kinks.
Bent hoses must be replaced.
The complete ilter system must be replaced in rotation after a maximum of ten years.
The hoses must be replaced in rotation after a maximum of ive years.
Caution: Before replacing ilter systems or hoses, read the technical data (Chapter 9) and the
!
operating and safety information (Chapter 3).
Regularly clean the outside of the ilter system with a soft, damp cloth.
Caution: Do not use any abrasive chemicals, cleaning solutions or astringent cleaning agents.
!
8 Troubleshooting
8.1 No water low
Cause:
Water intake closed.
Troubleshooting:
Open the water intake at the upstream stop valve or by closing the locking
handle
8.2 No or low water low in spite of open water intake
Cause:
Mains pressure too low.
Troubleshooting:
C heck mains pressure. If the fault persists, check the ilter system and ilter
cartridge and replace if necessary.
Cause:
Filter head not mounted in direction of low.
Troubleshooting:
Dismantle ilter head and install in direction of low (Chapter 4).
8.3 Leak
Cause:
Screwed connections not itted correctly.
Troubleshooting:
C heck mains pressure. Check all screwed connections and mount according to
Chapter 4. If the fault persists, replace ilter system.
on the ilter head
.
2
1
EN
21