KERN CFS Série Mode D'emploi page 151

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

14.3 Model KERN CFS 50K-3
 Switch-on balance and during the selftest press
 Use the numeric keys to enter password:
 Either input four times zero „0000" as a standard password,
or input the user-defined password (input cf. parameter
„Pin" chapter 13).
 Confirm input by
 Select bulk material or reference balance via
appeared [] displays the current balance.
When using as counting system, the bulk material balance
as well as the reference balance must be adjusted. The
adjustment process must be carried out on both balances.
 Acknowledge with
 Using
select the weighing unit [kg or lb] which will be used
for linearization. The appeared [] displays the current
weighing unit.
Acknowledge with
 Ensure that there are no objects on the weighing pan.
 Wait for stability display (indicator [] atop
appears), then press
 Carefully place required adjustment weight (chapter 1) into
the center of the weighing plate on display „LoAd".
 Wait for stability display, then press
 After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
In case of an adjustment error or incorrect adjusting weight
the display will show an error message (
adjustment process.
64
Operation
.
.
.
.
.
Display
„Pin"
.
„Pin" „----„
. The
„tECH" „LoCAL"
„tECH" „rEmotE"
„tECH" „Unit"
), repeat
CFS/CCS-BA-e-1624

Hide quick links:

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ccs série

Table des Matières