7 Basic Operation
7.1 Switching on/off
Actuate power switch (cf. chapter 2) at the right bottom of the scale to the front.
The balance will carry out a self-test As soon as the weight display appears, the
balance is ready for weighing.
To switch-off push backward the switch-on/switch-out on the right lower side of
the balance.
7.2 Zeroing
Resetting to zero corrects the influence of light soiling on the weighing plate. The
resetting range of the balance was adjusted to ± 2% max. by factory.
Further adjustments are possible in the menu, see chapter 12.
When using as counting system the zeroing range of both balances can be set in
the menu, see chapter 13.
Manual
Unload the balance
Press
, the balance starts resetting to zero.
The [] symbol over a appears.
Automatic
In the menu the automatic zero point correction can be switched off or the amount
can be changed, see chapter 13.
7.3 Switching over reference scale bulk scales if used as counting system
For piece counting a platform can be connected via the second balance interface.
In the counting system KERN CCS the quantity counting is made on the bulk
material balance KERN KFP. The reference balance KERN CFS permits due to its
high resolution a very precise determination of the average piece weight.
The second balance can be operated in the same manner like the first balance.
By pressing
the display changes from one to the other balance.
In the display appears
The appeared [] displays the active balance.
24
or
CFS/CCS-BA-e-1624
.