HomeFill® System
DANGER!
– This product is not intended as a life-sustaining or
life-supporting device.
– DO NOT SMOKE while using this device. Keep all
matches, lighted cigarettes or other sources of ignition
out of the room in which this product is located and
away from where oxygen is being delivered.
– NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning
can result in severe fire, property damage and cause
physical injury or death.
– The information contained in this document is subject
to change without notice.
Operating Information
Cylinder MUST be emptied of gas prior to shipping by courier or
postal service.
Only personnel instructed and trained in the use of the HomeFill
Conserving Cylinder should operate this product.
Product contains magnetic, ferrous material that may affect the
results of an MRI.
Be sure all connections are tight and leak free.
DO NOT use a liquid leak detector to test for leaks.
DO NOT autoclave.
DO NOT gas sterilize with ethylene oxide.
6
DO NOT clean with aromatic hydrocarbons, including, but not
limited to benzene and toluene.
DO NOT attempt to repair product.
DO NOT immerse product in any kind of liquid.
Store product in a clean area when not in use.
Only use Invacare Corporation carrying bag.
NEVER smoke in an area where oxygen is being administered.
ALWAYS confirm prescribed dose before administering to patient
and monitor on a frequent basis.
DO NOT use if dirt or contaminants are present on or around
cylinder, valve, product or connecting devices.
NEVER use near any type of flame or flammable/explosive substances,
vapours or atmosphere.
NO OXYGEN is delivered when the pointer is aligned with "O".
DO NOT block the outlet fitting or kink the cannula when the
product is in use, as this may damage the unit.
The use of this device is limited to the oxygen patient. Cylinders
MUST be used only by the oxygen patient and are not to be
distributed to any other individual for any purpose.
This product is NOT to be used by patients who breathe through
their mouths.
DO NOT allow oil from your hands or other sources to come
into contact with the compressor coupling or bottle fittings. These
substances may become flammable in the presence of oxygen and
cause injury. Avoid touching the bottle fitting whenever possible.
Example of items to avoid are hand creams or lotions, cooking oils,
suntan oils or sunblocks and similar oily products. If you use these
1176148-A