adequate pressure for accurate oxygen flow.
If needed, use an alternate source of oxygen
such as a flow control valve attached to the
Reservoir.
6. The Portable does not disengage easily from the
Reservoir after filling.
• The Portable and Reservoir fill connectors
may have become frozen. DO NOT USE
FORCE. Allow a few minutes for the frozen
parts to warm, then disengage the Portable
when the ice has melted. To prevent the units
from freezing together, always wipe the male
fill connector on the Reservoir and the female
fill connector on the Portable with a clean,
dry cloth before filling.
NOTE: It is recommended that routine maintenance
be performed annually on this equipment. Please
contact your Health Care Provider to ensure this
routine maintenance is being completed.
Cleaning Standard
WARNING: DO NOT ALLOW
WATER INTO ANY OF THE
CONTROLS, THE FILL CONNECTOR,
OR THE OXYGEN SUPPLY
CONNECTOR.
Caution: Clean only after the unit
is empty and vented.
• Clean using a solution of mild dish washing
detergent and water.
• Apply cleaning solution directly to a lint-free
cloth. Do not spray cleaners directly on the
H300/H850.
• Wipe the outside surface with the lint-free
cloth until the outside surface is clean.
Caution: Do not use high
temperature and high pressure
washing equipment to clean
these units.
• Do not get cleaner on any internal
components or valves.
• Allow the unit to dry thoroughly before
using.
HELiOS 300 & 850
WEEE and RoHS
This symbol is to remind the equipment
owners to return it to a recycling facility
at the end of its life, per Waste Electrical
and Electronic Equipment (WEEE)
Directive.
Our products will comply with the restriction of
Hazardous Substances (RoHS) directive. They will
not contain more than trace amounts of lead or
other hazardous material content.
Transport and Storage
The device should be stored in the upright position
well ventilated. Do not allow the device to be
tilted forward or laying on its front. Operating
temperature range of 14°F to 104°F (-10°C to
40°C). Storage temperature ranges from -40°F to
158°F (-40°C to 70°C).
Maintenance
Clean the fill connectors on both the stationary
and portable units with a clean, dry, lint-free cloth
between each fill to prevent freezing and possible
equipment failure.
NOTE: Any additional maintenance needed must
be done by a qualified service technician or service
provider.

Accessories

Backpack —HELiOS H850
Marathon
H850 Cart
PN 21033264-C4 C | User Manual
Hip pack —HELiOS H300
11 - ENG