  • FRANÇAIS, page 20
To change head - Fig. 9
To remove
When face, airway and jaw have been removed
(described in "Fig. 3"), flex the lower edges of the
head outward until the head can be removed.
To install
Grasp the manikin head with both hands, with
your fingers on the inside of the head. Gently pull
outwards and slide the head onto the neck posts,
one side at a time.
To install the airway, face and jaw, see page 4.
AED Little Anne only
To change battery
AED Little Anne comes with a standard alkaline
6LF22 / 6LR61 9 Volt battery.
- Open the chest skin.
Note: Be aware of wires connecting the chest
skin to the battery box under the rib plate.
- Remove the old battery and install a new
9 Volt alkaline battery. Be sure to follow the
positive (+) and negative (-) battery icons
etched into the inside of the battery holder
when inserting new battery.
Fig. 9
Battery low indicators
- Depressing the test button on the battery
board will illuminate the LEDs in the chest
skin if battery strength is sufficient for
operation. If the LEDs do not light when the
button is depressed a new battery is required.
- A "Battery low" voice prompt from the AED
Trainer 2 will be heard immediately after
connecting Laerdal Link training pads if
the manikin battery is low.