Commissioning, Operation; Mechanical Mounting - WIKA DG-10 Mode D'emploi

Masquer les pouces Voir aussi pour DG-10:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6. Commissioning, operation

6. Commissioning, operation
CAUTION!
GB
Prior to commissioning, the digital pressure gauge must be subjected to a visual inspection.
Only use the digital pressure gauge if it is in perfect condition with respect to safety.
Required tool: Open-ended spanner (spanner width 27)

6.1 Mechanical mounting

The sealing faces at the instrument have to be
undamaged and clean.
When screwing the instrument in, the force
required to do this must not be applied through
the case, but only through the spanner flats
provided for this purpose and using a suitable
tool.
The correct torque depends on the dimensions
of the process connection and the gasket used
(form/material).
Do not overwind the process connection, do not exceed the permissible torques.
When screwing in, do not cross the threads.
For information on tapped holes and welding sockets, see Technical information IN 00.14
at www.wika.com.
14
WIKA operating instructions digital pressure gauge, model DG-10
Spanner flats

Publicité

Table des Matières
loading

Table des Matières