6. Commissioning, operation
MIN/MAX memory
The MIN/MAX memory is reset to the current measured value with each measuring cycle.
Tare function "TARE" (model DG-10-E)
In this mode, the positive deviation of the measured value from the tared value (max. 20 % of the
measuring range) is shown on the main display. The non-tared value is shown on the bargraph display,
so that you can always read the current utilisation of the digital pressure gauge.
The following possibilities can be activated for the tare function:
Select the tare function via the "TARE" menu item. Use the arrow keys to adjust the value to be tared.
The adjusted value to be tared will be stored in the memory as tare.
Load the digital pressure gauge to the pressure to be tared. The main display will show the current
measured value. Via the menu, select the tare function, "TARE", and press the two arrow keys
simultaneously. The current measured value will be stored as tare. The main and secondary display
will be set to "0000".
The tare function can be reset in both variants by setting the tare value to "0000".