Specifications - Hach FL1500 Manuel D'utilisation De Base

Masquer les pouces Voir aussi pour FL1500:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 89
Table of Contents

Specifications

Specifications are subject to change without notice.
Specification
Dimensions (H x W x D)
Enclosure
Weight
Pollution degree
Installation category
Protection class
Power requirements
Backup battery option
Fuses
Storage conditions
Data storage
Communications
USB ports
Display
Sensor options
Sensor connections
Totalizer
Time-base accuracy
Operation
Details
25.4 × 29.2 × 12.1 cm (10.0 × 11.5 × 4.75 inch), enclosure with cover only
25.4 × 31.8 × 13.3 cm (10.0 × 12.5 × 5.25 inch), enclosure with cover and
mounting bracket
NEMA 4X, IP 66 (with and without removable cover)
3.2 kg (7.0 lb)
2
II
I
AC: 100-240 VAC, 50/60 Hz, 15 watts
DC: 10-30 VDC, 15 watts
12 VDC lead acid battery
AC power: two T 3.15 A, 250 VAC; DC power: one F 10 A, 250 VDC; Relays:
two T 3.15 A, 250 VAC
Temperature:
Logger only: -20 to 60 °C (-4 to 140 °F)
Logger and AC battery backup: -15 to 40 °C (5 to 104 °F)
0 to 95% non-condensing relative humidity
Altitude: 2000 m (6560 ft) maximum
-40 to 70 °C (-40 to 158 °F)
Maximum 829,440 measurements (180 days storage, 5-minute logging intervals
for 16 parameters); data wraps when full
USB and RS485 (Modbus)
USB A port for USB memory stick only, USB B port for PC only
QVGA, color
Flo-Dar, SVS, Flo-Tote 3, AV9000S (bare wire), US9001, US9003, BL9000,
digital differential pH, rain gauge
Basic model: 2; Advanced model: 4
Software totalizer, scalable
Connection for external mechanical totalizer (advanced model only)
Maximum 1 second per day
on page 22
on page 26
English 3

Publicité

Chapitres

Table des Matières
loading

Table des Matières